0577-62378177
lunte@lunte.com.cn
二维码
Product Center

Product Center

Lunte specializes in the design and manufacture of temperature sensors and nuclear measurement instruments, involved in industries such as industrial, nuclear power, and is recognized as a national-level "little giant" enterprise and a "hidden champion" enterprise in Zhejiang Province.

Home   <   Product Center   <   Othes
Temperature Transmitter
Temperature Transmitter
Temperature Transmitter
Temperature Transmitter

Temperature Transmitter

LT series temperature transmiter adopts HART bussing technique, which is a new generation of intelligent temperaturemodule and an indispensable field device in process control. The device integrates a weath offunctional modules, whichcan realize both general function detection and complex control stratecies. Wsing digital technology, it can be applied toa variety of RTD and thermocouple sensors. The interface between the field and the control room is simple, and the cosof installation, operation and maintenance can be greatly reduced.

分享

Product Overview

Temperature Module Size


Technical specifications of resistance temperature detectors (RTDs)

Technical specifications

Input signalPt 100, Pt 1000, Cu50, Cu100, resistance from 0 ~ 500Ω, 0 ~ 4000Ω  
Channel NumberSingle channel
RTD Connection2,3,4 wire system
Bus PowerHART bus 11~35 VDC
Bus SignalHART signal, current mode
Electrical isolation1000VAC
Operating temperature- 40℃~85℃
Humidity range5%~95%RH
Startup time≤ 5 seconds
Update time0.5 seconds
DisplayCOG128x64 dot matrix LCD
EMCComplies with GB/T 18268.1-2010 and EN 61326.1-2013
Protection GradeTerminal IP00, enclosure IP40
Voltage influence±0. 005%/V
Alarm signalAlarm output can be set to 3.5 to 23 mA  
Upper limit / lower limit current: 20.8 mA / 3.8 mA
Current accuracy0.03%
Intrinsic safety (optional)Ex ia IIC T4 Ga

Ambient temperature accuracy specifications

Signal typeRecommended usage rangeAccuracy (25°C)Temperature drift (per degree Celsius)
Pt100-200~850°C±0.15°C±0.003°C

Other technical specifications

Connection2、3、4
CMRR≥70dB (50 Hz And 60 Hz)
DMRR≥70dB (50 Hz And 60 Hz)

Thermocouple technical specifications

Normal temperature accuracy index

Signal typeRecommended use rangeAccuracy (25°C)Temperature (per ℃)
B500℃~1810℃±0.77℃±0.050℃
E-200℃~1000℃±0.20℃±0.025℃
J-190℃~1200℃±0.35℃±0.01℃
K-200℃~1372℃±0.40℃±0.025℃
N-190℃~1300℃±0.50℃±0.015℃
R0℃~1768℃±0.75℃±0.023℃
S0℃~1768℃±0.70℃±0.023℃
T-200℃~400℃±0.35℃±0.015℃

Other technical specifications

Compensation accuracy±0.5℃
Sensor typeB, E, J, N, K, R, S, T; -100mV ~+100mV Voltage
CMRR)≥70dB (50Hz And 60Hz)
DMRR≥70dB (50Hz And 60Hz)
Temperature drift<50ppm/℃


Recommended Products

Thermocouple for Furnace Regenerator Crown Thermocouple for Furnace Regenerator Crown

Thermocouple for Furnace Regenerator Crown

Used for temperature measurement at the top of coke oven regenerators, with a thermocouple insertion depth of nearly 2 meters. It features a reinforced square box and support ribs to prevent deformation when installed horizontally in high-temperature environments. The adjustable insertion depth is facilitated by a movable flange connection.

Connectors Connectors

Connectors

Flange Movable Thermocouple and RTD Flange Movable Thermocouple and RTD

Flange Movable Thermocouple and RTD

Assembly Thermocouple Assembly Thermocouple

Assembly Thermocouple

Aerospace Gasification Furnace Multipoint Thermocouple Aerospace Gasification Furnace Multipoint Thermocouple

Aerospace Gasification Furnace Multipoint Thermocouple

Flexible thermocouple Flexible thermocouple

Flexible thermocouple

Sheath of flexible thermocouple is drawn by special stainless steel in special process, so it is muchsofter thantraditional ones, but the performance is same, despite of flexibility. Flexible thermocouple can be inserted intozigzag thermowell and measure temperature of various structure inner the reactor.